Silex Medical, LLC recalls Signature Laparoscopic Instruments
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
Drawn from openFDA device enforcement reports. Filter by severity, status, state, year, or manufacturer.
Showing 25–36 of 39 recalls
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device.…
Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.
Medline ReNewal has identified that the drip chambers of affected devices may disconnect from the saline line tubing when under tension.
Medline ReNewal has identified that the drip chambers of affected devices may disconnect from the saline line tubing when under tension.
Medline ReNewal has identified that the drip chambers of affected devices may disconnect from the saline line tubing when under tension.
Medline ReNewal has identified that the drip chambers of affected devices may disconnect from the saline line tubing when under tension.
Precision Thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece.
Precision Thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece.
Some units in a single production lot were programmed incorrectly which will result in the VivAer Stylus being incorrectly recognized as RhinAer Stylus when connected to the Aerin Console.