Nova Biomedical Corporation recalls StatStrip Glucose and B- Ketone meter. Model No: 66429. Glucose test system.
A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) b…