Spacelabs Healthcare, Inc. recalls Brand Name: Spacelabs Ultraview SL Command Module Product Name: SL Command Module Mod…
Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).
Drawn from openFDA device enforcement reports. Filter by severity, status, state, year, or manufacturer.
Showing 1–7 of 7 recalls
Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO mea…
Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of anot…
Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of anot…
Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of anot…
Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring.…
Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring.…