Myofunctional Research Company USA recalls Myosa for Kids KS1 REF MyOSAKS1_Sb/Sp/Mb/Mp/Lb/Lc KS2 REF MyOSAKS2_Sb/Sc/Mb/Mc/Mp/L…
Due to firm mistakenly distributing product within the U.S. without FDA premarket clearance/approval.
Drawn from openFDA device enforcement reports. Filter by severity, status, state, year, or manufacturer.
Showing 133–144 of 439 recalls
Due to firm mistakenly distributing product within the U.S. without FDA premarket clearance/approval.
The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments.…
An instrument intended for sequencing DNA libraries with in vitro diagnostic assays is used with reagent kits including flow cells that have flow cell gaskets that have a quality issue, which may lead to flow cell leaks…
Test cuvettes, used with an assay for monitoring heparin anticoagulation, may have a foreign object that has sharp edges and could pose a risk of injury if not detected prior to handling.
Product testing did not meet expected stability criteria.
The IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures capa…
Due to Users not having access to the Instructions for Use/User Manual due to it being password protected.
Software error causing incorrect data to be displayed on the 24-hour Sensor Glucose Overview Graph, any potential therapy decisions were made based on the incorrect data displayed on the 24-hour sensor glucose overview g…
Potential that product was leak tested with equipment outside of its expected operating range, resulting in bronchoscopes that may not be susceptible to leaks.
Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.
Due to incorrect product label (Incorrect product name identified on outer packaging).
Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.