Baxter Healthcare Corporation recalls Brand Name: Welch Allyn
Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.
Drawn from openFDA device enforcement reports. Filter by severity, status, state, year, or manufacturer.
Showing 169–180 of 218 recalls
Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.
Reports of devices sparking/popping and potentially burning patients.
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and…
Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer.…
Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer.…
Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer.…
Under certain workflows, patient information shown in the viewer may not match the images displayed on Centricity Universal Viewer Zero Footprint Client (ZFP) with versions ZFP v6.0: SP11.6, SP11.6.0.1 and SP11.6.1 on sy…
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.