Siemens Medical Solutions USA, Inc recalls interventional fluoroscopic x-ray system
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails.…
Drawn from openFDA device enforcement reports. Filter by severity, status, state, year, or manufacturer.
Showing 37–48 of 50 recalls
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails.…
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails.…
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails.…
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails.…
The X-ray safety and manufacturing date labels affixed to the Lago X/ AMI HTX instrument do not correspond exactly to the labeling stipulated by the FDA / CDRH.
The X-ray safety and manufacturing date labels affixed to the Lago X/ AMI HTX instrument do not correspond exactly to the labeling stipulated by the FDA / CDRH.
The MSOT Acuity Echo does not include an adequate instruction for safe operation of the device. The previous version of the user manuals contained a wrong value for the Nominal Ocular Hazard Distance (NOHD).
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques.…
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques.…
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques.…
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques.…
A potential issue with ARTIS One systems was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration.…