Abiomed, Inc. recalls Impella RP Flex with SmartAssist. Product Code: 1000323.
Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.
Drawn from openFDA device enforcement reports. Filter by severity, status, state, year, or manufacturer.
Showing 97–108 of 371 recalls
Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.
Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.
A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.
The mix up of the impression caps provided in the package. The provided impression caps are magenta instead of brown.
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification.…
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification.…
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification.…
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification.…
Presence of particulates in affected devices that can be deposited into breast tissue during use.
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if…
Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new o…
Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new o…