Becton, Dickinson And Company recalls ForeSight Elite Tissue Oximetry Cable REF HEMFSM10 UDI-DI code: 00690103210927 The…
Due to incorrect switching of Channel 1 and 2 on the module and flag labels on the tissue oximetry cables.
Drawn from openFDA device enforcement reports. Filter by severity, status, state, year, or manufacturer.
Showing 1–10 of 10 recalls
Due to incorrect switching of Channel 1 and 2 on the module and flag labels on the tissue oximetry cables.
Incorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophageal and Tracheal Flange contains a Large Esophageal Flange.
Incorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophageal and Tracheal Flange contains a Large Esophageal Flange.
Some units of Aluminum Chloride Gel contain the package insert for Ferric Sulfate Gel. The incorrect insert misidentifies the product and contains different first aid instructions, most critically in that the Ferric Sulf…
Incorrect inner package labeling. Boxes labeled 4010-02-15-T3 markers may contain individual packages labeled as T4 Markers.
Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm.
Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm.
Mix up between specified lots of 2.5mm and 2.8mm outer diameter screws.
Mix up between specified lots of 2.5mm and 2.8mm outer diameter screws.
A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery.…