Augustine Temperature Management, LLC recalls

The FDA has posted 10 enforcement recalls from Augustine Temperature Management, LLC since 2025, none of them Class I (most serious). Most recent report: Feb 26, 2025.

Class II: 10
Class IIOngoingZ-1198-2025

Augustine Temperature Management, LLC recalls Hot Dog PATIENT WARMING

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of de…

  • Thermal Regulating System
  • Device Design
Augustine Temperature…MNFeb 26, 2025
Class IIOngoingZ-1197-2025

Augustine Temperature Management, LLC recalls Hot Dog PATIENT WARMING

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of de…

  • Thermal Regulating System
  • Device Design
Augustine Temperature…MNFeb 26, 2025
Class IIOngoingZ-1199-2025

Augustine Temperature Management, LLC recalls Hot Dog PATIENT WARMING

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of de…

  • Thermal Regulating System
  • Device Design
Augustine Temperature…MNFeb 26, 2025