The FDA has posted 35 enforcement recalls from Medtronic MiniMed, Inc. since 2025, none of them Class I (most serious). Most recent report: Jul 29, 2026.

Class II: 35
Class IIOngoingZ-2778-2026

Medtronic MiniMed, Inc. recalls MiniMed Go App

For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime.…

  • Drug Dose Calculator
  • Software design
Medtronic MiniMed, Inc…CAJul 29, 2026
Class IIOngoingZ-2698-2026

Medtronic MiniMed, Inc. recalls MiniMed 780G Insulin Pump

A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 follo…

  • Alternate Controller Enabled Insulin Infusion Pump
  • Software design
Medtronic MiniMed, Inc…CAJul 15, 2026
Class IIOngoingZ-1741-2026

Medtronic MiniMed, Inc. recalls MiniMed 780G Insulin Pump (MMT-1884

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infu…

  • Alternate Controller Enabled Insulin Infusion Pump
  • Device Design
Medtronic MiniMed, Inc…CAApr 8, 2026
Class IIOngoingZ-1748-2026

Medtronic MiniMed, Inc. recalls Paradigm Insulin Pump (MMT-712

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infu…

  • Insulin Infusion Pump
  • Device Design
Medtronic MiniMed, Inc…CAApr 8, 2026
Class IIOngoingZ-1740-2026

Medtronic MiniMed, Inc. recalls MiniMed 630G Insulin Pump (MMT-1714

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infu…

  • Threshold Suspend Automated Insulin Dosing
  • Device Design
Medtronic MiniMed, Inc…CAApr 8, 2026
Class IIOngoingZ-1751-2026

Medtronic MiniMed, Inc. recalls MiniMed 530G Insulin Pump (MMT-551

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infu…

  • Threshold Suspend Automated Insulin Dosing
  • Device Design
Medtronic MiniMed, Inc…CAApr 8, 2026
Class IIOngoingZ-1747-2026

Medtronic MiniMed, Inc. recalls MiniMed 620G Insulin Pump (MMT-1750)

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infu…

  • To Be Used With Invasive Glucose Sensor Insulin Infusion Pump
  • Device Design
Medtronic MiniMed, Inc…CAApr 8, 2026
Class IIOngoingZ-1743-2026

Medtronic MiniMed, Inc. recalls MiniMed 640G Insulin Pump (MMT-1711

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infu…

  • To Be Used With Invasive Glucose Sensor Insulin Infusion Pump
  • Device Design
Medtronic MiniMed, Inc…CAApr 8, 2026
Class IIOngoingZ-1749-2026

Medtronic MiniMed, Inc. recalls Paradigm REAL-Time Insulin Pump (MMT-522

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infu…

  • To Be Used With Invasive Glucose Sensor Insulin Infusion Pump
  • Device Design
Medtronic MiniMed, Inc…CAApr 8, 2026
Class IIOngoingZ-1742-2026

Medtronic MiniMed, Inc. recalls MiniMed 770G Insulin Pump (MMT-1880

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infu…

  • Single Hormonal Control Automated Insulin Dosing Device System
  • Device Design
Medtronic MiniMed, Inc…CAApr 8, 2026
Class IIOngoingZ-1746-2026

Medtronic MiniMed, Inc. recalls MiniMed 740G Insulin Pump (MMT-1811

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infu…

  • To Be Used With Invasive Glucose Sensor Insulin Infusion Pump
  • Device Design
Medtronic MiniMed, Inc…CAApr 8, 2026
Class IIOngoingZ-1745-2026

Medtronic MiniMed, Inc. recalls MiniMed 720G Insulin Pump (MMT-1809

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infu…

  • To Be Used With Invasive Glucose Sensor Insulin Infusion Pump
  • Device Design
Medtronic MiniMed, Inc…CAApr 8, 2026