mo-Vis BVBA recalls

The FDA has posted 6 enforcement recalls from mo-Vis BVBA since 2025, including 6 Class I (most serious) recalls. Most recent report: Sep 10, 2025.

Class I: 6
Class IOngoingZ-2456-2025

Mo-Vis BVBA recalls IDM-MICRO-R. Electrical wheelchair component.

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power…

  • Wheelchair Components
  • Device Design
mo-Vis BVBASep 10, 2025
Class IOngoingZ-2454-2025

Mo-Vis BVBA recalls Micro Joystick R-net. Electrical wheelchair component.

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power…

  • Wheelchair Components
  • Device Design
mo-Vis BVBASep 10, 2025
Class IOngoingZ-2459-2025

Mo-Vis BVBA recalls IDM-ARLITE-R. Electrical wheelchair component.

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power…

  • Wheelchair Components
  • Device Design
mo-Vis BVBASep 10, 2025
Class IOngoingZ-2457-2025

Mo-Vis BVBA recalls IDM-MULTI-R. Electrical wheelchair component.

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power…

  • Wheelchair Components
  • Device Design
mo-Vis BVBASep 10, 2025
Class IOngoingZ-2455-2025

Mo-Vis BVBA recalls Multi Joystick R-net. Electrical wheelchair component.

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power…

  • Wheelchair Components
  • Device Design
mo-Vis BVBASep 10, 2025