Medline Industries, LP recalls Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED
Reason for recall
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED, MANUALMEDLITEUDI-DI 00080196288284
What the firm is doing
Firm initially notified customers November 26, 2025. On August 14, 2026, the firm sent a follow-up letter which provides additional product SKU - Item numbers for the electrical system safety risks and additional important information related to the previous notices. The notice explained the issue, potential risks, and requested the user ONLY USE MEDLINE SIDE RAILS and MEDLINE ACCESSORIES that are intended for use with Medline Homecare Beds. When using Medline Side Rails and Medline Accessories, follow all applicable instructions for use. Distributors were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 orrecalls@medline.com.
DistributionShow detailsHide
Distribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama.
A reasonable probability that use of the product will cause serious adverse health consequences or death.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-3228-2026
- FDA device classification · FNJOfficial FDA classification for this device type
- CFR regulation · 21 CFR 880.5120The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Medline Industries, LPSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class I recalls
Draeger, Inc. recalls VentStar Resus Neo. Neonatal. 25 Units per Box. Part Number: MP00310 Disposable brea…
Potential cracks in the hose, which could lead to leakage or the hose falling off.
- Positive End Expiratory Pressure Breathing Attachment
- Under Investigation by firm
Medline Industries, LP recalls Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire.…
- Adjustable Home-Use Ac-Powered Therapeutic Bed
- Device Design
Boston Scientific Corporation recalls Imager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5
Tip may fracture or detach, leading to device fragments remaining within the patient or migrating through the vasculature. Patient impact may include prolonged procedure or life-threatening embolism.
- Diagnostic Intravascular Catheter
- Process control