Draeger, Inc. recalls VentStar Resus Neo. Neonatal. 25 Units per Box. Part Number: MP00310 Disposable brea…
Reason for recall
Potential cracks in the hose, which could lead to leakage or the hose falling off.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
VentStar Resus Neo. Neonatal. 25 Units per Box. Part Number: MP00310 Disposable breathing circuit for use with a resuscitation module.
Lot / code information
- Part / Item #
- MP00310. UDIDI: 04048675422358. All
What the firm is doing
Draeger notified consignees on about 09/03/2026 via emailed Urgent Medical Device Recall letter. Consignees were informed of the issue of a cracked hose and that currently there are no alternative products from other manufacturers that have been tested and approved by Draeger for use with these devices, nor are there substitute products available within the Draeger portfolio in the U.S. Consignees were instructed to carefully inspect the hoses for damages prior to use according to the product s Instructions for Use (IFU), discard products that show cracks or have a detached hose, inform all potential users within the facility, and if affected products have been transferred or distributed, to notify those customers. Additionally, customers were requested to complete and return the provided Medical Device Recall Return Response Acknowledgement and Receipt Form. Draeger will replace the affected circuits free of charge after a correction is implemented based on the quantity indicated on the Medical Device Recall Return Response Acknowledgement and Receipt Form.
DistributionShow detailsHide
US Nationwide distribution including in the states of IN, AK, AL, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SD, TN, TX, VT, WA, WI, WV, WY, MS, SC, UT, VA, AR, and territories AS and PR. The countries of Canada, Germany, United Kingdom, Spain, Italy, Ireland, Poland, Australia, Switzerland, Netherlands, Portugal, France, Austria, Hong Kong, Malaysia, Costa Rica, Andorra, Belgium, Sweden, Cyprus, Greece, Latvia, Romania, Albania, Moldavia, Kazakhstan, Croatia, Libya, Mauritius, Honduras, Suriname, Uruguay, Syria, Israel, Kuwait, Bahrain, Qatar, Utd.Arab.Emir., Oman, Denmark, Norway, New Zealand, Brunei Darussalam, Singapore, China, India, Japan.
A reasonable probability that use of the product will cause serious adverse health consequences or death.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-3252-2026
- FDA 510(k) clearance · K092029The device's official FDA premarket clearance record
- FDA device classification · BYEOfficial FDA classification for this device type
- CFR regulation · 21 CFR 868.5965The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Draeger, Inc.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class I recalls
Medline Industries, LP recalls Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire.…
- Manual Bed
- Device Design
Medline Industries, LP recalls Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire.…
- Adjustable Home-Use Ac-Powered Therapeutic Bed
- Device Design
Boston Scientific Corporation recalls Imager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5
Tip may fracture or detach, leading to device fragments remaining within the patient or migrating through the vasculature. Patient impact may include prolonged procedure or life-threatening embolism.
- Diagnostic Intravascular Catheter
- Process control