SunMed Holdings, LLC recalls Brand Name: Broselow Product Name: Broselow Domestic Complete ALS Organizer Model/Cat…
Reason for recall
Product contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Brand Name: Broselow Product Name: Broselow Domestic Complete ALS Organizer Model/Catalog Number: 7730ALS Software Version: Not Applicable Product Description: The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care. Component: NoUDI-DI Each-108894835891514 affected lots000431228400043145030004333110Each-10889483589151
What the firm is doing
AirLife issued an URGENT: MEDICAL DEVICE RECALL NOTICE to its consignees on 12/15/2025 via email and US mail. The notice explained the issue, potential risk to the patient, and requested the following: "Customer immediate actions: Please take the following actions immediately: 1. Stop using the affected tapes o Immediately discontinue use of all AirLife Broselow Rainbow Tapes identified as 2025 Edition, 36-23446 Rev 3 Print Version. 2. Identify and segregate affected product o Examine your inventory and clinical areas for affected Broselow Rainbow Tapes (see Image B for identifying characteristics). o Remove all affected tapes from clinical service. o Segregate or quarantine the affected tapes to prevent further use. 3. Follow disposition instructions o Follow the instructions provided in the accompanying communication regarding discarding/destruction of the affected tapes. o Do not redistribute or place any affected tapes back into service. 4. Notify your internal users o Ensure that all clinicians and healthcare professionals within your organization who may use the Broselow Rainbow Tape are informed of this recall/field removal and understand that the affected Rev 3 tapes must not be used. 5. Notify downstream customers (if applicable) o If you have further distributed the affected tapes, please identify your customers/consignees and notify them of this recall/field removal promptly. o Your notification may be enhanced by including a copy of this letter. o If you have impacted product on hand, do not ship it; instead, hold it for discarding/destruction as instructed. 6. Complete and return the Response Form o Please complete and return the attached Response Form via e-mail to productquality@myairlife.com as soon as possible. Please complete and return Attachment B if you have affected product for discarding/destruction. This allows us to document your receipt of this recall notice and the status of affected product at
DistributionShow detailsHide
Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Singapore.
A reasonable probability that use of the product will cause serious adverse health consequences or death.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-1116-2026
- FDA device classification · OKIOfficial FDA classification for this device type
- CFR regulation · 21 CFR 878.4040The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find SunMed Holdings, LLCSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class I recalls
Spectra Medical Devices, LLC recalls Brand Name: Elevaris/Spectra Medical Devices Product Name: Sodium Chloride Injection
Manufacturing firm, Huons, was cited for issues regarding product quality and sterility assurance. If the product sterility is compromised, patients may be exposed to microorganisms, endotoxins, or foreign matter.…
- Vascular Access Flush Saline
- Nonconforming Material/Component
Baxter Healthcare Corporation recalls Baxter Duo-Vent Solution Set and Male Luer Lock Adapter
Baxter received customer reports of air bubbles in the drip chamber and tubing when a pressure cuff is in use or when the tubing is flushed at the fully open position.
- Intravascular Administration Set
- Under Investigation by firm
Boston Scientific Neuromodulation Corporation recalls INFINION CX LEAD KIT
Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, i…
- Totally Implanted For Pain Relief Spinal-Cord Stimulator
- Device Design