ARROW INTERNATIONAL, LLC recalls Description/REF: ACCESS TRAY/GU-04020
Reason for recall
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Description/1 code
- ACCESS TRAY/GU-04020
PSI KIT: 8.5 FR/AK-09803-CDC
PSI KIT: 9 FR/AK-09903-CDC
PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-09810-PCMH1
PSI KIT: 9 FR X 10 CM/ASK-09903-KH
PSI KIT: 9 FR X 10 CM/ASK-09903-PHP1
PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-UPA
PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-YNH
PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09907-SL
PSI KIT: 8.5 FR X 4-1/8IN (10 CM)/NA-09803-CDC
PSI KIT: 9 FR X 4-1/8IN (10 CM)/NA-09903-CDC
PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-ECL4
PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-MGH3
MULTI-LUMEN/PSI KIT/AK-11142-SPCS
MAC KIT: 2-L 9 FR DISTAL X 11.5 CM/CDC-11242-1A
MULTI-LUMEN/PSI KIT/AK-11142-D
MULTILUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-11142-MGH4
MULTI-LUMEN/PSI KIT/NA-11242-CDC
MULTI-LUMEN/PSI KIT/AK-21142-SK
MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-LMDTG
MULTI LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-RH
MULTILUMEN/PSI KIT 9 FR X 11.5 CM/ASK-21242-IMC
MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/ASK-21242-LMDTG
MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMD
MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UU3
MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-VCU
MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-1A
MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-X1A
MULTI-LUMEN/PSI KIT/AK-21242-CDC
2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-FH4
MULTI LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-GH1
2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-SUU
2-L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-UPMC4
MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UKL
MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMF
MULTILUMEN/PSI KIT: 9 FR X 4 1/2 IN/ASK-21242-UR1
PSI KIT: 8.5 FR/AK-29803-CDC
PSI KIT: 9 FR/AK-29903-CDC
PSI KIT: 9 FR/AK-29907-CDC
PSI KIT: 8.5 FR X 4IN (10 CM)/ASK-09804-UN1
PSI KIT: 9 FR/ASK-09903-UM
PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-LMDT
PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-LMDTG
PSI KIT: 9 FR X 4 1/8 (10 CM)/ASK-29903-BID1
PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-LMDTG
PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-UMD
PSI KIT: 9 FR X 4 IN (10 CM)/ASK-29903-UR1
PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/CDC-29803-X1A
PSI KIT: 8.5 FR X 10 CM/ASK-29803-ECL
PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-UPM3
PSI KIT: 9 FR X 10 CM/ASK-29903-ECL
PSI KIT: 9FR X 4IN (10 CM) ANTIMICROBIAL/ASK-29903-UPM3
PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09907-TG1
Lot / code information
- REF
- (Material)/UDI-DI/
- Lot #
(Batch): GU-04020/30801902118807, 10801902193275, 30801902118807, 30801902118807/33F24C0753, 33F24E0736, 33F24G0154 — +469 moreShow all
(Batch): GU-04020/30801902118807, 10801902193275, 30801902118807, 30801902118807/33F24C0753, 33F24E0736, 33F24G0154, 33F24G0648, 33F24H0377, 33F25A0059, 33F25A0072, 33F25A0395, 33F25F0260, 33F25G0736, 33F25H0298, 33F25J0603, 33F26B0084, 33F26B0314, 33F26C0578, 33F24D0036; AK-09803-CDC/20801902121756, 10801902193275, 10801902117417/33F24B0126, 33F24B0127, 33F24C0021, 33F24D0021, 33F24E0072, 33F24G0338, 33F24G0587, 33F24J0506, 33F24K0105, 33F24L0722, 33F25A0468, 33F25C0107, 33F25C0953, 33F25E0054, 33F25E0921, 33F25G0086, 33F25H0023, 33F25J0828, 33F25K0255, 33F25K0364, 33F25L0328, 33F24A0239; AK-09903-CDC/10801902117417, 10801902193275, 20801902121916/33F24A0356, 33F24B0128, 33F24B0129,33F24C0022, 33F24D0022, 33F24E0074, 33F24E0075, 33F24G0340, 33F24G0588, 33F24J0507, 33F24K0107, 33F24L0676, 33F24L0759, 33F25A0470, 33F25C0108, 33F25C0730, 33F25E0056, 33F25E0923, 33F25H0025, 33F25J0471, 33F25K0031, 33F25L0237, 33F25M0920, 33F26A1053, 33F24A0386; ASK-09810-PCMH1/10801902193275/33F25K0051, 33F24J0109; ASK-09903-KH/10801902119350, 10801902193275/33F24D0713, 33F24K0385, 33F24L1060, 33F25D0340, 33F25E0778, 33F25H0567, 33F25M0163, 33F26A1407, 33F25E1184; ASK-09903-PHP1/10801902193275, 10801902115222/33F24J0115, 33F25H0497, 33F26A0410, 33F24F0625; ASK-09903-UPA/10801902193275, 10801902118193/33F24K0387, 33F24L0855, 33F24A0787; ASK-09903-YNH/10801902118193, 10801902193275, 10801902118209/33F24E0527, 33F24H0478, 33F24L0682, 33F25H0135, 33F25J0915, 33F24A0110; ASK-09907-SL/10801902118209, 10801902193275/33F24D0053, 33F24D0967, 33F24G0265, 33F25E0215, 33F25E0454, 33F25G0680, 33F25J0551, 33F25K0328, 33F25L0206, 33F26C0312, 33F25B0188; NA-09803-CDC/10801902119930, 10801902193275, 10801902119947/33F24C0756, 33F24E0739, 33F25A0342, 33F25J0615, 33F26B0028, 33F24A1327; NA-09903-CDC/10801902119947, 10801902193275, 10801902119954/33F24E0740, 33F24H0381, 33F24J0199, 33F24J0361, 33F24L0020, 33F25B0278, 33F25C0551, 33F25E0340, 33F25E0400, 33F25E0468, 33F25F0419, 33F25H0708, 33F26A1655, 33F26B0023, 33F24A0113; ASK-09903-ECL4/10801902193275, 10801902224344/33F26A1582, 33F26B0399, 33F25E0534; ASK-09903-MGH3/10801902193275, 10801902189841, 10801902212006/33F25C0264, 33F25G0338, 33F25J0360, 33F25K0130, 33F26B0126, 33F26C0004, 33F24D0780; AK-11142-SPCS/10801902196412, 10801902193275, 60801902123093/33F24A0382, 33F24A1019, 33F24C0024, 33F24E0078, 33F24E0154, 33F24G0343, 33F24J0510, 33F24K0110, 33F24L0673, 33F25A0472, 33F25C0381, 33F25C0389, 33F25C0952, 33F25E0059, 33F25H0027, 33F25J0831, 33F25K0302, 33F25L0365, 33F26B0514, 33S24J0510, 33F24D0075; CDC-11242-1A/60801902116736, 10801902193275, 40801902111768/33F24D0431, 33F24E0213, 33F24F0140, 33F24F0220, 33F24G0103, 33F24H0392, 33F24H0425, 33F24H0450, 33F24L0011, 33F25B0018, 33F25B0096, 33F25D0705, 33F25E1120, 33F25F0024, 33F25F0414, 33F25G0043, 33F25G0486, 33F25J0016, 33F25J0992, 33F25K0308, 33F25M0481, 33F24C1001; AK-11142-D/20801902121916, 10801902193275, 10801902196412/33F24A1018, 33F24C0023,33F24E0077, 33F24G0342,33F24J0509, 33F24K0109, 33F24L0615, 33F25A0471, 33F25C0842, 33F25E0058, 33F25G0088, 33F25J0472, 33F25K0301, 33F25L0467, 33F26B0593, 33F26B0958, 33F24A0252; ASK-11142-MGH4/10801902193275, 10801902189865, 10801902182545/33F25D0466, 33F25F0113, 33F25L0130, 33F25M0398, 33F26A1539, 33F26B0680, 33F24D0736; NA-11242-CDC/10801902119954, 10801902193275, 10801902119978/33F24A0817, 33F24A1218, 33F24B0399, 33F24C0270, 33F24G0554, 33F24J0663, 33F24L0743, 33F25C0197, 33F25C0870, 33F25E0273, 33F25H0179, 33F25H0761, 33F25L0419, 33F26B0562, 33F24C0271; AK-21142-SK/10801902193275, 30801902122248, 30801902122255/33F24G0356, 33F24G0589, 33F24J0520, 33F24K0122, 33F24L0670, 33F25A0480, 33F25C0720, 33F25E0071, 33F25H0030, 33F25J0837, 33F25L0473, 33F25M0964, 33F26A1097, 33F26B0956, 33R24L0670, 33F24D0023; ASK-21142-LMDTG/10801902193275/33F25K0417, 33F25F0007; ASK-21142-RH/10801902193275/10801902129953/33F24J0108, 33F24L0165, 33F24M0172, 33F25B0422, 33F25E1172, 33F25H0776, 33F25K0424, 33F26C0337, 33F25A0245; ASK-21242-IMC/10801902193275, 10801902118407/33F24G0500, 33F24J0631, 33F24K0255, 33F24L0609, 33F25A0572, 33F25E0223, 33F25H0142, 33F25J0921, 33F26B0543, 33F25C0696; ASK-21242-LMDTG/10801902193275, 10801902220827/33F25G0202, 33F25H0414, 33F25K0382, 33F25M0130, 33F26C0378, 33F24H0123; ASK-21242-UMD/10801902193275/33F24K0476, 33F24L0315, 33F25B0196, 33F25C0279, 33F25D0476, 33F25H0492, 33F25J0378, 33F25M0150, 33F26A1365, 33F25B0551; ASK-21242-UU3/10801902193275/33F24G0503, 33F24L0652, 33F25C0181, 33F25E0226, 33F25H0144, 33F25J0556, 33F24K0077; ASK-21242-VCU/10801902193275, 10801902112962/33F24K0694, 33F25G0171, 33F25H0495, 33F25J0379, 33F26C0494, 33F25C0439; CDC-21242-1A/50801902116821, 10801902193275/33F24D0486, 33F24D0981, 33F24E0227, 33F24F0175, 33F24G0104, 33F24H0296, 33F24H0305, 33F24H0331, 33F24H0357, 33F24H0393, 33F24H0427, 33F24M0019, 33F24M0055, 33F25A0131, 33F25A0367, 33F25B0019, 33F25C0511, 33F25D0706, 33F25E0417, 33F25E1122, 33F25F0416, 33F25G0066, 33F25G0476, 33F25G0711, 33F25H0852, 33F25J0964, 33F25K0352, 33F25L0229, 33F26A1614, 33F26A1624, 33F26A1649, 33F24H0332; CDC-21242-X1A/10801902193275, 30801902116834, 10801902196511/33F25J0291, 33F25J0672, 33F25J0678, 33F25J0690, 33F25J0700, 33F25J0708, 33F25J0724, 33F25L0067, 33F26A1616, 33F26A1625, 33F26B0642, 36F25B0014, 36F25E0020, 36F25G0029, 36F25H0016, 36F25J0005, 33F24C0518; AK-21242-CDC/30801902122255, 10801902193275, 30801902122262/33F24E0082, 33F24E0083, 33F24J0521, 33F24K0123, 33F24L0618, 33F25A0481, 33F25C0946, 33F25E0072, 33F25H0031, 33F25K0037, 33F25K0414, 33F25L0372, 33F26B0520, 33F26C0626, 33F24E0352; ASK-21142-FH4/10801902193275/33F24L0683; ASK-21142-GH1/10801902193275/33F25C0180, 33F25E1023, 33F25H0141, 33F25H0752, 33F25J0920, 33F25L0455, 33F25B0276; ASK-21142-SUU/10801902193275/33F25J0096, 33F25C0275; ASK-21142-UPMC4/10801902193275/33F25A0247; ASK-21242-UKL/10801902114065, 10801902193275, 10801902136487/33F24F0710, 33F24K0467, 33F24L0314, 33F25B0195, 33F25C0278, 33F25D0475, 33 …
What the firm is doing
On 4/21/2026, correction notices (Firm Event Number: EIF-000608, lidocaine and bupivacaine) were mailed to customers who were asked to do the following: Arrow International LLC is voluntarily recalling product due to the Huons Co. Ltd. recall of Lidocaine and Bupivacaine, which is included in impacted Arrow Kits & Sets. Medical facilities 1. Provide this correction notice to all persons who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 2. Check your inventory for product within the scope of this Medical Device Notification and place a copy of this notice with all affected product. 3. Complete and return the response form virtually via: https://fca.myteleflex.com/en/recall/000608 Distributors: 1. Provide Medical Device Notification to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor. 2. Check your inventory for affected product. Place a copy of this notification with all affected product and ensure that the notification is included with future distributions of affected product. 3. Complete and return the response form via link: https://fca.myteleflex.com/en/recall/000608 4. If you have further distributed product outside of your country, please notify Teleflex Customer Service by return email. If you have questions contact firm's Customer Service: Link: https://fca.myteleflex.com/en/recall/000608, Customer Service: Telephone: 1-866-396-2111, FAX: 1-855-419-8507 Email: Recalls@teleflex.com On 5/29/2026, additional amended notices (Firm Event Number: EIF-000608-01) were mailed to include additional affected devices. On 5/1/2026, additional notices (Firm Event Number: EIF-000609, sodium chloride) were mailed and emailed to customers who have 0.9% sodium chloride (10mL saline solution ampule) in kits and sets.
DistributionShow detailsHide
US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
A reasonable probability that use of the product will cause serious adverse health consequences or death.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2385-2026
- FDA 510(k) clearance · K002507The device's official FDA premarket clearance record
- FDA 510(k) clearance · K011761The device's official FDA premarket clearance record
- FDA 510(k) clearance · K780532The device's official FDA premarket clearance record
- FDA 510(k) clearance · K940079The device's official FDA premarket clearance record
- FDA device classification · DYBOfficial FDA classification for this device type
- CFR regulation · 21 CFR 870.1340The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find ARROW INTERNATIONAL, LLCSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class I recalls
Draeger, Inc. recalls VentStar Resus Neo. Neonatal. 25 Units per Box. Part Number: MP00310 Disposable brea…
Potential cracks in the hose, which could lead to leakage or the hose falling off.
- Positive End Expiratory Pressure Breathing Attachment
- Under Investigation by firm
Medline Industries, LP recalls Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire.…
- Manual Bed
- Device Design
Medline Industries, LP recalls Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire.…
- Adjustable Home-Use Ac-Powered Therapeutic Bed
- Device Design