Skip to content
RecallWatchMedical Device Safety
Class IOngoingZ-2679-2026

Oscor Inc. recalls Abiomed REF: 0052-3021

Oscor Inc.Palm Harbor, FL, United StatesReported Jul 22, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Abiomed REF: 0052-3021, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F, 23F. heart pump and accessories
    GTIN 00885672009786/00813502010558

What the firm is doing

On June 10, 2026, Oscor (an Integer subsidiary ) issued a Urgent Medical Device Recall notification to affected consignees. Oscor asked consignees to take the following actions: 1. Product removal is not required, and hospitals may continue to use existing inventory. 2. Physicians should continue to use appropriate clinical awareness, monitoring, and patient support as needed. 3. Consider a device exchange or using the repositioning sheath to minimize blood loss if potential leakage from the sidearm and under the sheath cap and along the hub score lines in the hub region is observed. 4.Review the information within this letter, complete all fields, sign, and return the attached business response form (BRF) (refer to Attachment 1) to Abiomed3958@sedgwick.com. 5. Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). 6.If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice. 7. Post a copy of this notice in a visible area for awareness.

DistributionShow details

Distribution in the US state of MA

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-2709-2026

Argon Medical Devices, Inc recalls Bone Marrow Tray

Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.

  • Biopsy Instrument
  • Material/Component Contamination
Argon Medical Devices,…TXJul 22, 2026
Class IOngoingZ-2708-2026

Argon Medical Devices, Inc recalls Arterial Line Kit 20ga x 6

Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.

  • X-Ray Detectable Internal Gauze/Sponge
  • Material/Component Contamination
Argon Medical Devices,…TXJul 22, 2026
Class IOngoingZ-2707-2026

Argon Medical Devices, Inc recalls T-Lok Bone Marrow Tray

Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.

  • Biopsy Instrument
  • Material/Component Contamination
Argon Medical Devices,…TXJul 22, 2026