CooperSurgical, Inc. recalls INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP
Reason for recall
Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2709, 44-2710, 44-271244-2715Model Number19 affected lots55001972244-270955003161755003378155003162144-2710550031615550031616
+11 more
55003162055003476044-271255003161155003161455003377955003161944-2715550031618550033780550034763
What the firm is doing
CooperSurgical notified consignees on about 08/11/2026 via "URGENT Medical Device Recall" letter. Consignees were notified of the issue and potential hazard, and were instructed to identify affected units on hand, if any loose connection is recognized during the use of the CPAP, replace the tubing with another tubing or CPAP device, and remove all product of the subject lots from inventory and note the number of units to be returned on the attached Recall Response Form that is provided. Distributors were instructed to remove all product of the subject lots from inventory and note the number of units to be returned on the attached Distributor Only Acknowledgement Form. Additionally, they were instructed to communicate the recall notification to all customers that they distributed these devices to.
DistributionShow detailsHide
Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.
A reasonable probability that use of the product will cause serious adverse health consequences or death.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-3139-2026
- FDA 510(k) clearance · K911780The device's official FDA premarket clearance record
- FDA device classification · BZDOfficial FDA classification for this device type
- CFR regulation · 21 CFR 868.5905The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find CooperSurgical, Inc.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class I recalls
Boston Scientific Corporation recalls Imager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5
Tip may fracture or detach, leading to device fragments remaining within the patient or migrating through the vasculature. Patient impact may include prolonged procedure or life-threatening embolism.
- Diagnostic Intravascular Catheter
- Process control
AVID Medical, Inc. recalls Halyard IV START KIT NON-STERILE
Convenience kits containing BD ChloraPrep Clear - 1 mL Applicator, which may be contaminated with Aspergillus penicillioides.
- General Surgery Tray
- Packaging process control
CooperSurgical, Inc. recalls INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP
Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and…
- Non-Continuous (Respirator) Ventilator
- Device Design