AVID Medical, Inc. recalls Halyard IV START KIT NON-STERILE

AVID Medical, Inc.Toano, VA, United StatesReported Sep 30, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

Convenience kits containing BD ChloraPrep Clear - 1 mL Applicator, which may be contaminated with Aspergillus penicillioides.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Halyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit Model Number: KRIV26-02.

Lot / code information

Model
KRIV26-02
UDI
10809160460481
Lot #
1658437. Expiration Date: 1/31/2027

What the firm is doing

Consignees were initially notified on about 07/22/2026 via "URGENT: MEDICAL DEVICE FIELD CORRECTION" letter. They were informed that product may exhibit an open or incomplete seal on the packaging of the applicator. Consignees were instructed to quarantine impacted product, discontinue use of the BD ChloraPrep Clear - 1 mL Applicators within the kits, add a warning label to instruct users to discard the impacted component, and notify customers if the kits had been further distributed. Consignees were notified again on about 08/20/2026 with an updated "URGENT: MEDICAL DEVICE FIELD CORRECTION" letter. Consignees were informed that certain lots of the BD ChloraPrep Applicators are contaminated with Aspergillus penicillioides. Instructions included to discard impacted kits containing affected applicators and notify customers if the kits had been further distributed.

DistributionShow details

US Nationwide distribution.

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-3140-2026

CooperSurgical, Inc. recalls INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP

Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and…

  • Non-Continuous (Respirator) Ventilator
  • Device Design
CooperSurgical, Inc.CTSep 30, 2026
Class IOngoingZ-3139-2026

CooperSurgical, Inc. recalls INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP

Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and…

  • Non-Continuous (Respirator) Ventilator
  • Device Design
CooperSurgical, Inc.CTSep 30, 2026