CooperSurgical, Inc. recalls INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP

CooperSurgical, Inc.Trumbull, CT, United StatesReported Sep 30, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-070944-0715
    Model Number
    7 affected lots
    55003162444-070755002105744-070955003162344-0715550031625

What the firm is doing

CooperSurgical notified consignees on about 08/11/2026 via "URGENT Medical Device Recall" letter. Consignees were notified of the issue and potential hazard, and were instructed to identify affected units on hand, if any loose connection is recognized during the use of the CPAP, replace the tubing with another tubing or CPAP device, and remove all product of the subject lots from inventory and note the number of units to be returned on the attached Recall Response Form that is provided. Distributors were instructed to remove all product of the subject lots from inventory and note the number of units to be returned on the attached Distributor Only Acknowledgement Form. Additionally, they were instructed to communicate the recall notification to all customers that they distributed these devices to.

DistributionShow details

Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-3139-2026

CooperSurgical, Inc. recalls INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP

Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and…

  • Non-Continuous (Respirator) Ventilator
  • Device Design
CooperSurgical, Inc.CTSep 30, 2026